Objective To analyze the electrical resistance of a newly developed neuromonitoring pedicle screw (Neuro-PS) and to verify the electrophysiologic properties of the Neuro-PS in a pig model.
Methods We developed 2 types of the Neuro-PS in which a gold lead was located internally (type I) and externally (type II). We measured the electrical resistance of the Neuro-PS and the conventional screw and analyzed the electrical thresholds of triggered EMG (t-EMG) of each screw by intentionally penetrating the medial pedicle wall and contacting the exiting nerve root in a pig model.
Results The electrical resistances of the Neuro-PS were remarkably lower than that of the conventional screw. In electrophysiologic testing, only the type II Neuro-PS under the leadnerve contact condition showed a significantly lower stimulation threshold as compared to the conventional screw.
Conclusion The Neuro-PS demonstrated lower electrical resistances than the conventional screw. The type II Neuro-PS under the lead-nerve contact condition showed a significantly lower stimulation threshold compared to that of the other screws in the t-EMG test.
Citations
Citations to this article as recorded by
Feasibility of Using Intraoperative Neurophysiological Monitoring for Detecting Bone Layer of Cervical Spine Surgery Weiyang Zuo, Lingjia Yu, Haining Tan, Xiang Li, Bin Zhu, Yuquan Liu, Xuan Peng, Yong Yang, Qi Fei Clinical Spine Surgery.2024; 37(10): E480. CrossRef
Intraoperative triggered electromyographic monitoring of pedicle screw efficiently reduces the lumbar pedicle breach and re-operative rate-a retrospective analysis based on postoperative computed tomography scan Tong Yongjun, Zhao Yuntian, Chen Biao, Jiang Zenghui BMC Musculoskeletal Disorders.2023;[Epub] CrossRef