Development of Neuromonitoring Pedicle Screw - Results of Electrical Resistance and Neurophysiologic Test in Pig Model

Neurospine 2021;18(1):117-125.
Published online: November 18, 2020

1Department of Neurosurgery, Topspine Hospital, Daegu, Korea

2Department of Electrical Engineering, Kyungpook National University, Daegu, Korea

3Department of Neurosurgery, Kyungpook National University Hospital, School of Medicine, Kyungpook National University, Daegu, Korea

4Department of Orthopedic surgery, Columbia University Medical Center/New York Presbyterian Hospital, New York, NY, USA

5Department of Rehabilitation Medicine, Chung-Ang University College of Medicine, Seoul, Korea

6Department of Neurosurgery, CHA Bundang Medical Center, CHA University, Seongnam, Korea

7Department of Neurosurgery, Seoul National University Bundang Hospital, Seongnam, Korea

8Department of Neurosurgery, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea

9Department of Neurosurgery, Chung-Ang University College of Medicine, Seoul, Korea

Corresponding Author Kyoung-Tae Kim https://orcid.org/0000-0003-4867-6854 Department of Neurosurgery, Kyungpook National University Hospital, 130 Dongdeok-ro, Jung-gu, Daegu 41944, Korea Email: nskimkt7@gmail.com

Woo-Seok Bang and Jonghoo Park contributed equally to this study as cofirst authors.

• Received: July 17, 2020   • Revised: September 3, 2020   • Accepted: September 3, 2020

Copyright © 2021 by the Korean Spinal Neurosurgery Society

This is an open access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

  • 12,752 Views
  • 271 Download
  • 2 Web of Science
  • 2 Crossref
  • 2 Scopus
prev next

Citations

Citations to this article as recorded by  Crossref logo
  • Feasibility of Using Intraoperative Neurophysiological Monitoring for Detecting Bone Layer of Cervical Spine Surgery
    Weiyang Zuo, Lingjia Yu, Haining Tan, Xiang Li, Bin Zhu, Yuquan Liu, Xuan Peng, Yong Yang, Qi Fei
    Clinical Spine Surgery.2024; 37(10): E480.     CrossRef
  • Intraoperative triggered electromyographic monitoring of pedicle screw efficiently reduces the lumbar pedicle breach and re-operative rate-a retrospective analysis based on postoperative computed tomography scan
    Tong Yongjun, Zhao Yuntian, Chen Biao, Jiang Zenghui
    BMC Musculoskeletal Disorders.2023;[Epub]     CrossRef

Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:

Include:

Development of Neuromonitoring Pedicle Screw - Results of Electrical Resistance and Neurophysiologic Test in Pig Model
Neurospine. 2021;18(1):117-125.   Published online November 18, 2020
Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:
Include:
Development of Neuromonitoring Pedicle Screw - Results of Electrical Resistance and Neurophysiologic Test in Pig Model
Neurospine. 2021;18(1):117-125.   Published online November 18, 2020
Close

Figure

  • 0
  • 1
  • 2
  • 3
  • 4
Development of Neuromonitoring Pedicle Screw - Results of Electrical Resistance and Neurophysiologic Test in Pig Model
Image Image Image Image Image
Fig. 1. Neuromonitoring pedicle screw (Neuro-PS): Manufactured pedicle screws. (A, B) The type I Neuro-PS was constructed by inserting a 0.5-mm diameter gold lead through the center of a conventional screw constructed of Ti-6Al-4V. (C, D) The type II Neuro-PS was constructed by placing a 0.5-mm diameter gold lead along the surface of a conventional screw constructed of Ti-6Al-4V. The white arrow indicates the gold lead placed at the center of the type I Neuro-PS.
Fig. 2. Photograph of the screw penetration of the medial wall in a pig model. The screws penetrated the medial pedicle wall, and made contact with the exiting roots. The white arrows indicate the exiting root at 5 and 6 levels of the lumbar spine.
Fig. 3. Diagram of the 2 different conditions of the type II neuromonitoring pedicle screw (Neuro-PS). (A) The type II Neuro-PS under the lead-nerve contact condition. (B) The type II Neuro-PS under the lead-nerve noncontact condition.
Fig. 4. Simulated and measured resistances of the control, the type I and the type II neuromonitoring pedicle screw (Neuro-PS). (A) The bars without fill patterns show the measured resistance in air of the control, the type I and the type II Neuro-PS for the screw size of 6.5 mm × 40 mm. The bars with fill patterns represent the resistance calculated by COMSOL Multiphysics simulation. (B) Simulated voltage distribution across the control, the type I and the type II Neuro-PS when a 1 ampere current is applied to one end (head) of the gold lead while the other end (tip) is connected to the ground. The maximum voltage indicates the resistance of the screw.
Fig. 5. The volume charge density distribution of the neuromonitoring pedicle screw (Neuro-PS) calculated by COMSOL Multiphysics simulation software. (A, B) The volume charge density distribution of type I Neuro-PS in a cross-section and midsagittal section of the screw, respectively. (C, D) The volume charge density distribution of type II Neuro-PS in a cross-section and midsagittal section of the screw, respectively. The unit of the volume charge density is Coulomb per cubic meter (C/m3).
Development of Neuromonitoring Pedicle Screw - Results of Electrical Resistance and Neurophysiologic Test in Pig Model
Screw size (mm) (diameter-length) No. Type Average resistance in air (mΩ) Average resistance in bone (mΩ)
4.5 × 40 5 Control 96 ± 7.08 95 ± 6.98
5 Type I 5.4 ± 0.51 (p < 0.001)* 5.4 ± 0.45 (p < 0.001)* (p = 0.894)**
5 Type II 3.2 ± 0.61 (p < 0.001)* 2.9 ± 0.31 (p < 0.001)* (p = 0.405)**
6.5 × 40 5 Control 91 ± 4.98 89 ± 4.81
5 Type I 5.4 ± 0.43 (p < 0.001)* 5.4 ± 0.58 (p < 0.001)* (p = 0.913)**
5 Type II 3.1 ± 0.43 (p < 0.001)* 2.6 ± 0.41 (p < 0.001)* (p = 0.098)**
6.5 × 50 5 Control 102 ± 6.38 99 ± 6.31
5 Type I 5.5 ± 0.42 (p < 0.001)* 5.4 ± 0.38 (p < 0.001)* (p = 0.904)**
5 Type II 3.3 ± 0.52 (p < 0.001)* 2.9 ± 0.55 (p < 0.001)* (p = 0.513)**
Nerve root Control screw Type I screw Type II screw
Type II screw
Under the lead nerve contact condition Under the lead nerve noncontact condition
L3 root 7.50 ± 0.50 mA 7.00 ± 0 mA (p = 0.343)* 3.00 ± 0.71 mA (p = 0.029)* (p = 0.029)** 7.00 ± 0.71 mA (p = 0.486)*
L4 root 5.75 ± 0.43 mA 5.25 ± 0.43 mA (p = 0.343)* 2.75 ± 0.43 mA (p = 0.029)* (p = 0.029)** 5.75 ± 0.43 mA (p > 0.99)*
L5 root 6.50 ± 0.87 mA 6.75 ± 0.43 mA (p = 0.886)* 2.75 ± 0.43 mA (p = 0.029)* (p = 0.029)** 6.75 ± 0.43 mA (p = 0.886)*
L6 root 7.00 ± 0.70 mA 6.50 ± 0.5 mA (p = 0.486)* 3.25 ± 0.43 mA (p = 0.029)* (p = 0.029)** 6.75 ± 0.43 mA (p = 0.687)*
Total average 6.69 ± 0.94 mA 6.38 ± 0.8 mA (p = 0.376)* 2.94 ± 0.57 mA (p < 0.001)* (p < 0.001)** 6.56 ± 0.73 mA (p = 0.637)*
Table 1. Electrical resistance comparison of 3 different pedicle screws in air and bone

Values are presented as mean±standard deviation.

p-value compared with control group.

p-value compared with “in air” group.

Table 2. Mean stimulation thresholds of screws at L3, L4, L5, and L6 root in pig model

Values are presented as mean±standard deviation.

p-value means compared with control group.

p-value means compared with control group.

p-value means compared with type II screw under the lead nerve noncontact condition.