Biomechanical Impact of Cement Augmentation on Pedicle Screw Fixation and Adjacent Segment Disease in Multilevel Lumbar Fusion: A Finite Element Analysis

Neurospine 2025;22(3):763-773.
Published online: September 30, 2025

1Department of Biomedical Engineering, Inje University, Gimhae, Korea

2GS Medical Co. Ltd., Cheongju, Korea

3Department of Orthopedic Surgery, Seoul St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea

4Department of Orthopedic Surgery, Eunpyeong St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea

Corresponding Author Hyung-Youl Park Department of Orthopedic Surgery, Eunpyeong St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, 1021 Tongil-ro, Eunpyeong-gu, Seoul 03321, Korea Email: matrixbest@naver.com
• Received: March 6, 2025   • Revised: May 14, 2025   • Accepted: May 18, 2025

Copyright © 2025 by the Korean Spinal Neurosurgery Society

This is an open access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

  • 6,699 Views
  • 85 Download
  • 4 Web of Science
  • 3 Crossref
  • 4 Scopus
prev next

Citations

Citations to this article as recorded by  Crossref logo
  • Enhancing Predictive Accuracy in Finite Element Analysis of Cement Augmentation: Methodological Considerations – A Commentary on “Biomechanical Impact of Cement Augmentation on Pedicle Screw Fixation and Adjacent Segment Disease in Multilevel Lumbar Fusio
    Fan Mo, Shaoqi He
    Neurospine.2026; 23(2): 500.     CrossRef
  • Reply Letter: A Commentary on “Biomechanical Impact of Cement Augmentation on Pedicle Screw Fixation and Adjacent Segment Disease in Multilevel Lumbar Fusion: A Finite Element Analysis”
    Hyung-Youl Park
    Neurospine.2026; 23(2): 502.     CrossRef
  • Biomechanical advances in oblique lateral lumbar interbody fusion using different interbody cages
    Yermike Bieshanbieke, Fei Xie, Baofeng Yan
    Journal of Orthopaedic Surgery and Research.2026;[Epub]     CrossRef

Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:

Include:

Biomechanical Impact of Cement Augmentation on Pedicle Screw Fixation and Adjacent Segment Disease in Multilevel Lumbar Fusion: A Finite Element Analysis
Neurospine. 2025;22(3):763-773.   Published online September 30, 2025
Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:
Include:
Biomechanical Impact of Cement Augmentation on Pedicle Screw Fixation and Adjacent Segment Disease in Multilevel Lumbar Fusion: A Finite Element Analysis
Neurospine. 2025;22(3):763-773.   Published online September 30, 2025
Close

Figure

  • 0
  • 1
  • 2
  • 3
  • 4
Biomechanical Impact of Cement Augmentation on Pedicle Screw Fixation and Adjacent Segment Disease in Multilevel Lumbar Fusion: A Finite Element Analysis
Image Image Image Image Image
Fig. 1. (A) Cement-augmented pedicle screw insertion at the upper instrumented vertebra (L2) and cement augmentation (vertebroplasty) at the L1 body. (B) Six types of samples. Type 1, without cement augmentation; type 2, L2 cement-augmented screws; type 3, L2 cement-augmented screws with L1 vertebroplasty.
Fig. 2. Biomechanical analyses. (A) Screw pull-out at L2. (B) Range of motion of the L1–2 segment. (C) Intradiscal pressure at the L1–2 level. (D) Screw breakage at L2. (E) Fracture risk at L1. PVMS, peak von Mises stress.
Fig. 3. Pull-out rates of L2 screws. Osteoporosis significantly increased the risk of screw pull-out across all stabilization types, with type 1 being the most highly affected. Type 1, without cement augmentation; type 2, L2 cement-augmented screws; type 3, L2 cement-augmented screws with L1 vertebroplasty; type 1-OP, type 1 in osteoporotic vertebra; type 2-OP, type 2 in osteoporotic vertebra; type 3-OP, type 3 in osteoporotic vertebra.
Fig. 4. Risk of adjacent segment disease. (A) Range of motion at L1–2. (B) Intradiscal pressure at L1–2. Type 1, without cement augmentation; type 2, L2 cement-augmented screws; type 3, L2 cement-augmented screws with L1 vertebroplasty; type 1-OP, type 1 in osteoporotic vertebra; type 2-OP, type 2 in osteoporotic vertebra; type 3-OP, type 3 in osteoporotic vertebra.
Fig. 5. Risk of implant failure. (A) Failure of L2 pedicle screws. (B) Failure of the L1 vertebral body (lower endplate). Type 1, without cement augmentation; type 2, L2 cement-augmented screws; type 3, L2 cement-augmented screws with L1 vertebroplasty; type 1-OP, type 1 in osteoporotic vertebra; type 2-OP, type 2 in osteoporotic vertebra; type 3-OP, type 3 in osteoporotic vertebra.
Biomechanical Impact of Cement Augmentation on Pedicle Screw Fixation and Adjacent Segment Disease in Multilevel Lumbar Fusion: A Finite Element Analysis
Materials Young modulus E (MPa)
Poisson ratio
Normal Osteoporosis Normal Osteoporosis
Bony structure
 Cancellous bone 150 50 0.4 0.4
 End plate 100 20 0.4 0.4
 Annulus ground 4.2 6 0.45 0.45
 Nucleus pulposus 1.0 81 0.499 0.3
 Cortical bone 12,000 0.3
 Posterior element 3,500 0.4
Implant
 PEEK (cage) 3,500 0.3
 Ti6A14V (pedicle screw) 110,000 0.34
 PMMA (bone cement) 3,500 0.3
Spinal movement Type 1 Type 2 Type 3 Type 1-OP Type 2-OP Type 3-OP
Failure of pedicle screw at L2
 Flexion PVMS 112.31 132.33 132.11 206.40 220.28 227.80
Failure 13.88 15.37 15.34 24.00 25.61 26.49
 Extension PVMS 110.60 124.70 120.90 189.10 208.70 203.80
Failure 13.54 14.51 14.14 21.98 24.27 25.69
 Left axial rotation PVMS 296.80 305.40 304.90 620.00 622.70 672.30
Failure 34.51 35.51 35.45 72.09 72.41 78.17
 Right axial rotation PVMS 301.70 311.80 310.70 657.60 673.20 699.90
Failure 35.08 36.25 36.19 76.47 78.28 81.38
 Left lateral bending PVMS 149.60 159.50 159.00 314.30 351.90 378.10
Failure 17.39 18.54 18.48 36.55 40.92 43.96
 Right lateral bending PVMS 145.10 157.30 156.50 320.40 350.50 377.30
Failure 16.87 18.29 18.19 37.26 40.76 43.87
Failure of L1 vertebral body
 Flexion PVMS 7.30 7.42 7.55 8.61 8.64 12.40
Failure rate 44.76 45.55 46.32 52.82 53.01 76.07
 Extension PVMS 6.87 6.90 7.04 7.72 7.74 10.92
Failure 42.15 42.33 41.16 47.35 47.48 66.99
 Left axial rotation PVMS 2.18 2.25 2.25 2.57 2.86 5.92
Failure 13.37 13.80 13.80 15.77 17.55 36.32
 Right axial rotation PVMS 2.26 2.32 2.32 2.67 2.78 6.47
Failure 13.85 14.23 14.23 16.38 17.06 39.69
 Left lateral bending PVMS 5.92 6.12 6.18 8.32 8.93 13.35
Failure 36.29 37.55 37.91 51.04 54.79 81.90
 Right lateral bending PVMS 5.66 5.82 5.88 8.61 9.23 13.48
Failure 34.72 35.71 36.07 52.82 56.63 82.70
Table 1. Material properties of the implant and spine used in finite element models

PEEK, polyetheretherketone; PMMA, polymethylmethacrylate.

Table 2. Failure analysis of L2 pedicle screw and L1 vertebral body

The unit of PVMS is MPa, and failure values are expressed as probabilities in percentage (%).

OP, osteoporosis; PVMS, peak von Mises stress.

Type 1, without cement augmentation; type 2, L2 cement-augmented screws; type 3, L2 cement-augmented screws with L1 vertebroplasty; type 1-OP, type 1 in osteoporotic vertebra; type 2-OP, type 2 in osteoporotic vertebra; type 3-OP, type 3 in osteoporotic vertebra.